Audits SaMD compliance to 21 CFR Part 820, IEC 62304, ISO 13485, ISO 14971. Reviews DHFs, technical files, validation protocols, and CAPAs with severity-rated findings.
From antigravity-awesome-skillsnpx claudepluginhub sickn33/antigravity-awesome-skills --plugin antigravity-awesome-skillsThis skill uses the workspace's default tool permissions.
Designs and optimizes AI agent action spaces, tool definitions, observation formats, error recovery, and context for higher task completion rates.
Enables AI agents to execute x402 payments with per-task budgets, spending controls, and non-custodial wallets via MCP tools. Use when agents pay for APIs, services, or other agents.
Compares coding agents like Claude Code and Aider on custom YAML-defined codebase tasks using git worktrees, measuring pass rate, cost, time, and consistency.
This skill transforms your AI assistant into a specialized MedTech Compliance Auditor. It focuses on Software as a Medical Device (SaMD) and traditional medical equipment regulations, including 21 CFR Part 820 (Quality System Regulation), IEC 62304 (Software Lifecycle), ISO 13485, and ISO 14971 (Risk Management).
@fda-medtech-compliance-auditor and provide the document you wish to review.Scenario: A CAPA was opened for a software defect in a Class II device. The documented root cause is “developer error — unclear requirements.” The corrective action is developer retraining.
Finding:
FDA AUDIT FINDING
Severity: Major
Citation: 21 CFR 820.100(a)(2) / IEC 62304 Section 5.1
Analysis:
"Developer error" is a symptom, not a root cause. Retraining alone is
a known red flag for FDA inspectors and will not withstand scrutiny.
The true root cause lies in the software requirements engineering
process itself — not an individual.
Required Actions:
1. Perform a 5-Whys or Fishbone analysis targeting the requirements
gathering and review process.
2. Update the SRS (Software Requirements Specification) and the
corresponding process SOP.
3. Document an effectiveness check with a measurable criterion
(e.g., zero requirements-related defects in next 3 releases).
4. Do not close the CAPA on retraining alone.